78 Maintaining your Library
For AIT libraries, the possible values are 0 to 96.
For DLT/SDLT libraries, the possible values are 0 to 60.
For LTO libraries, the possible values are 0 to 72.
Starting slot — indicates the starting slot. This field is read-only and only changes when New start
slot is selected.
New start slot — indicates whether you want to change the start slot or use the currently indicated
slot. Possible values are Yes and No.
3 Select OK.
4 With New start slot selected, press OK.
5 If you did not choose to modify the start slot, the exercise runs. If you chose to select a new start slot,
do the following:
a. With Select Column selected, press Change Column until the arrow points to the start
column. Press Next.
b. With Magazine selected, select Change Magazine until the arrow points to the start magazine.
Press Next.
c. With Slot selected, press Change Slot until the arrow points to the start slot.
The Select SOURCE line displays the start slot. For example, the display below has slot 2A4 selected
as the source slot.
d. Select OK. The exercise is run.
6 When the exercise is complete, a report displays.
Opening/Closing and Extending/Retracting
Use Step to manipulate the open/close and reach/retract features of the picker fingers.
Path: Main Menu > More > Service > Library > Diags > Fingers > Step