Service Action Codes 6-19
September 2007
31 00 00 A fatal error is detected in SCSI port 2.
Action 1
Verify that the SCSI bus connected to Library Port 2 is properly terminated
and powered.
Action 2
Verify that the SCSI ID is unique for the SCSI cable.
Action 3
Verify that the Library SCSI Port 2 is properly configured by using the
operator panel (Main Menu Setup Library SCSI Ports).
Action 4
If the problem is not corrected or persists, contact GCH. For GCH contact
information, refer to Getting More Information or Help on page 1-8.
32 00 00 Wrong SCSI bus connection is detected. A Single Ended SCSI bus is
connected to the Library Port 1 Differential Adapter.
Action 1
Verify that the Host SCSI bus is the same type as the Library SCSI Adapter
card in Port 1.
Action 2
If the problem is not corrected or persists, contact GCH. For GCH contact
information, refer to Getting More Information or Help on page 1-8.
33 00 00 Wrong SCSI bus connection is detected. A Single Ended SCSI bus is
connected to the Library Port 2 Differential Adapter.
Action 1
Verify that the Host SCSI bus is the same type as the Library SCSI Adapter
card in Port 2.
Action 2
If the problem is not corrected or persists, contact GCH. For GCH contact
information, refer to Getting More Information or Help on page 1-8.
34 00 00 A general SCSI failure is detected.
Action 1
Verify that the SCSI host is working properly.
Action 2
If the problem is not corrected or persists, contact GCH. For GCH contact
information, refer to Getting More Information or Help on page 1-8.
Table 6-2 Service Action Codes (Continued)
Reported
SAC
Perform these actions