Quantum Scalar i2000 Guide de l'utilisateur Page 423

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 544
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 422
Chapter 6 Maintaining Your Library
Maintenance Actions
Scalar i6000 User’s Guide 410
Running Calibration Teach 6
Starting the calibration teach process causes the library to assess the
position and alignment of various library components through the use of
calibration targets. Use this process to avoid cartridge-handling problems
caused by rack, drive, or I/E station misalignments.
Rack alignment calibration targets are tabs that are located on two special
magazines in each d
rive-side and door-side storage rack. I/E station
targets are small square holes that are located at the top and bottom of the
I/E station. Whenever you perform work on the library that could affect
the position of rack, drive, or I/E station calibration targets, even slightly,
you should initiate the calibration teach process.
When the library reaches 20,000 moves after the
last calibration occurred, and if then the library is
rebooted or an access door is closed, the library
automatically recalibrates itself.
1 Log on as an administrator.
2 Ma
ke sure that you are viewing the physical library. From the View
menu, click the name of the physical library.
3 Click Tool
s > Teach.
The Teach dialog box appears with Configure selected by default.
Note
If the physical library is not offline, you receive a
message that asks you whether you want to take it
offline. Click Yes.
Vue de la page 422
1 2 ... 418 419 420 421 422 423 424 425 426 427 428 ... 543 544

Commentaires sur ces manuels

Pas de commentaire